5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

505

Cancel determination
The determination is canceled and the next determination of the series
is started. If an exit track is present, it will be called. No error track will
be called.
Cancel determination and series
Both the determination and the series are stopped. If an exit track is
present, it will not be called. An error track will be called.

Parameters

Shift rate

Selection of the rate at which the rack rotates.

Input range

3 - 20 °/s

Default value

20 °/s

Shift direction

Selection of the shift direction.

Selection

auto | + |

Default value

auto

auto
The shift direction with the smallest rotation angle is selected automat-
ically.
+
Rack rotates counterclockwise.

Rack rotates clockwise.

Swing rate

Selects the rate of the robotic arm when moving to a rack position or a
special beaker (only with multiple row racks).

Input range

10 - 55 °/s

Default value

55 °/s

5.5.2.8.3

SWING

5.5.2.8.3.1

SWING - Overview

Dialog window: Method

SWING Properties... SWING - 'Command name'

Command for swinging the robotic arm. This command cannot be
executed unless the Sample Processor is equipped with a 786 Swing
Head
with robotic arm.