5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

480

■■■■■■■■

viva 1.0 (for Process analysis)

Addition volume #

Size of the individual volume that is to be dosed in. The number of addi-
tion volumes is defined by the amount defined in Number. If the com-
mand is called more often than volumes were defined, then the last vol-
ume is added in each case.

Input range

0.010 - 1,000.000 mL (Increment: 0.001)

Default value

0.100 mL

Addition
Selection of the addition mode for dosing standard solutions.

Selection

Add manually | Already added | Add with dos-
ing device

Default value

Add manually

Add manually
Manual addition of the solution, at which time the track is stopped. A
message appears and an acoustic signal is heard.
Already added
The solution is located already in the measuring vessel or was chan-
neled into the measuring vessel, e.g. with a peristaltic pump. The track
is not stopped. No message appears and no acoustic signal is heard.
Add with dosing device
Automatic addition of the solution with a dosing device (possible only
with solutions from the solution table). The solution is added through
the Dosing port 1 defined in the configuration for the dosing unit
(see Chapter 6.6.3.2, page 699).

Message
With manual addition, the running of the track in which the command is
located is stopped and a message is displayed.

Only editable for Addition = Add manually, otherwise disabled (includ-
ing text input field).

Selection

Display standard message | Display message
defined by the user

Default value

Display standard message

Display standard message
If this check box is activated, then the standard message for additions
will be displayed.
Display message defined by the user
If this check box is activated, then the message defined in the text field
will be displayed.