5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

481

Dosing device
Parameters for automatic dosing. This section is displayed only when
Addition = Add with dosing device.

Dosing rate

Rate (volume per time unit) of expulsion. The maximum rate depends on
the cylinder volume of the buret used and on the diameter of the hose at
the selected dosing port.

Input range

0.01 - 166.00 mL/min

Default value

2.00 mL/min

Selection

maximum

CAUTION

The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillary tubes with a diameter < 1 mm.

Filling rate

Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the buret used and on the diameter of the hose at the
selected fill port.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

5.5.2.7.4

ADD SAMPLE

5.5.2.7.4.1

ADD SAMPLE - Overview

Dialog window: Method

ADD AUX Properties... ADD SAMPLE - 'Command

name'

Command for the addition of the sample.

Appearance
The command has the following appearance:

Parameters
The parameters for the ADD SAMPLE command are configured in the
following dialog window: