4 cvs - sweep – Metrohm viva 1.0 (process analysis) User Manual
Page 419
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
407
Selection
off
Default value
off
Waiting time 2
Waiting time during which the Potential 2 potential is applied to the
electrodes.
Only editable for Potential 2
≠ off, otherwise inactive.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s
Potential 3
Potential that is applied to the electrodes during the Waiting time 3.
Input range
-5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
Waiting time 3
Waiting time during which the Potential 3 potential is applied to the
electrodes.
Only editable for Potential 3
≠ off, otherwise inactive.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s
Equilibration time
Waiting time before the sweep during which the Start potential is
applied to the electrodes.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
5.0 s
5.5.2.3.2.4
CVS - Sweep
Tab: Method
▶ CVS ▶ Properties... ▶ Sweep
Command name
Name of the command.
Entry
25 characters
Start potential
Start potential for the sweep.
Input range
-5.000 - 5.000 V (Increment: 0.001)
Default value
1.625 V