3 track, Track – Metrohm viva 1.0 (process analysis) User Manual

Page 402

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

390

■■■■■■■■

viva 1.0 (for Process analysis)

Action

NOTE

Is only displayed for sample data variables of the Number or Date/
Time
type.

Action

on | off (Default value: off)

If this option is enabled, then one of the two following actions will auto-
matically be executed if the limit values are breached.

Selection

Cancel determination | Cancel determination
and series

Default value

Cancel determination

Cancel determination
If this option is selected, then the ongoing determination will be can-
celed (without database entry). The next determination of the series
will then be started. If an exit track is present, it will be executed. If an
error track is present, it will not be executed.
Cancel determination and series
If this option is selected, then the ongoing determination will be can-
celed (without database entry). The next determination of the series
will not be started. If an exit track is present, it will be executed. If an
error track is present, it will not be executed.

5.5.2.2.3

TRACK

Dialog window: Method

TRACK Properties... TRACK - 'Command name'

Start command for normal track.

Appearance
The command has the following appearance:

Command name

Name of the command.

Entry

1 - 25 characters

Default value

Track %1

Advertising