Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1000

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

984

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.6.6, page 987).

[Delete]

Delete the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is activated, then the associated time and temperature for
the minimum measured value will be interpolated from the measuring
point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.5 ... 10.0 µA/s

Default value

5.0 µA/s

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is activated, then the associated time and temperature for
the maximum measured value will be interpolated from the measuring
point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.5 ... 10.0 µA/s

Default value

5.0 µA/s

Break point evaluation

Note

This evaluation method is possible only with Titrando, 855 and 867.

Advertising