3 digital in - stop criteria – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1312

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1296

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

5.6.11.3.3

DIGITAL IN - Stop criteria

Tab: Method

DIGITAL IN Stop criteria

Command name

Name of the command.

Input

25 characters

Stop criteria
Definition of possible stop criteria. The first stop criterion to be fulfilled will
stop the command.

Maximum run time

on | off (Default value: off)

Check box for switching on/off the stop criterion Maximum run time. If
the maximum run time is reached, then the command will be stopped.

Range

0 ... 999.9

Default value

0

Selection

min | s

Default value

min

Note

The run time is interrupted if the method is stopped with [HOLD].

Condition

on | off (Default value: off)

Check box for switching on/off the stop criterion Condition.

Expression for the condition(s) which can be created or edited with the

formula editor after pressing on

or double-clicking in the text field

(see Chapter 2.4, page 32).

The condition is monitored on a continuous basis; if it is fulfilled (evalua-
tion of the formula gives the result 1 = true), then the command will be
interrupted.

Input

1000 characters

Default value

empty