5.6.4.19.3
STDADD auto - Standard addition
Tab: Method
▶ STDADD ▶ Properties... ▶ Standard addition
Command name
Name of the command.
Input
25 characters
The parameters for the standard addition can be defined on this tab
Number of additions
Number of times the standard solution is added.
Range
1 ... 19
Default value
3
Volume auxiliary solution
Volume of the auxiliary solution (e.g. buffer) which was added to the sam-
ple before the first standard addition.
Range
0.0000 ... 9999.999 mL
Default value
0.0000 mL
Stop volume
Stops when the sum of the volume increments added exceeds the stop
volume.
Range
0.0 ... 9999.9 mL
Default value
100.0 mL
Addition solution
Note
Displayed only for Solution = not defined.
Concentration
Concentration of addition solution.
Range
0.001 ... 999999.999
Default value
1.000
Device
Concentration unit of addition solution.