Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 606

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
590
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Window = Measured value
Range
-20.000 ... 20.000 pH
Default value
-20.000 pH
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Window = Measured value pH
Range
-20.0 ... 20.0 pH
Default value
20.0 pH
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
Window = Time
Range
0 ... 999999 s
Default value
999999 s
Gran evaluation
Note
This evaluation method is possible only with Titrando and 855.
Gran evaluation
on | off (Default value: off)
If this option is enabled, then titration curves will be evaluated in accord-
ance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 930).
Procedure
Selection of the Gran procedure.
Selection
Normalized | Standard
Default value
Normalized