7 set ipol - conditioning, Conditioning – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 771

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
755
Selection
off
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.
Titrando, 855
Range
0 ... 999999 s
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.4.5.7
SET Ipol - Conditioning
Tab: Method
▶ SET Ipol ▶ Properties... ▶ Conditioning
Command name
Name of the command.
Input
25 characters
Conditioning
on | off (Default value: off)
Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.