5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

596

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Parameters
The parameters for the command MET U are configured in the following
7 tabs:

■

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

■

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

■

Titration parameters
Parameters for the run of the titration.

■

Stop conditions
Definitions of conditions which cause the stop of the titration.

■

Potentiometric evaluation
Parameters for the potentiometric evaluation of titration curves.

■

Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.

■

Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.

Command variables
The following command variables are generated in the method run by the
command MET U and can be used in formulas under the designation
'Command name.Variable designation':

Designation

Description

.BLV

Blank value of the sensor used for the command (only
for ISE sensors)

.BP{x}.DME

Measured value differential for the break point × (1…
9)

.BP{x}.MEA

Measured value for the break point × (1…9) in the unit
of the measured value

.BP{x}.TEM

Temperature for the break point × (1…9) in °C

.BP{x}.TIM

Time for the break point × (1…9) in s

.BP{x}.VOL

Volume for the break point × (1…9) in mL

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started