Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1425

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1409

used (see below). When the function is carried out the speed will auto-
matically be reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Dosing rate Dosing port 2

Speed at which dosing is to be carried out at Dosing port 2. The maxi-
mum dosing rate depends on the cylinder volume of the exchange unit
used (see below). When the function is carried out the speed will auto-
matically be reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Dosing rate Fill port

Speed at which dosing or filling is to be carried out at the Fill port. The
maximum dosing rate depends on the cylinder volume of the exchange
unit used (see below). When the function is carried out the speed will
automatically be reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Dosing rate Special port

Speed at which dosing is to be carried out at the special port. The maxi-
mum dosing rate depends on the cylinder volume of the exchange unit
used (see below). When the function is carried out the speed will auto-
matically be reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

The maximum dosing rate depends on the cylinder volume of the
dosing unit used (see below):

Cylinder volume

Maximum rate

2 mL

6.67 mL/min

5 mL

16.67 mL/min

10 mL

33.3 mL/min

20 mL

66.6 mL/min

50 mL

166.0 mL/min