5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

461

5.6.3.2.2

DET pH

5.6.3.2.2.1

DET pH - Overview

Dialog window: Method

▶ DET pH ▶ Properties.. ▶ DET pH - 'Command name'

Command for Dynamic equivalence point titrations with potentio-
metric pH measurement.

Instruments
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the command DET pH are configured on the follow-
ing 7 tabs:

■

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

■

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

■

Titration parameters
Parameters for the run of the titration.

■

Stop conditions
Definitions of conditions which cause the stop of the titration.

■

Potentiometric evaluation
Parameters for the potentiometric evaluation of titration curves.

■

Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.

■

Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.