4 cal spec - calibration wavelengths, 5 cal spec - calibration wavelength – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1123

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1107

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.

Range

1 ... 10000

Default value

10

5.6.5.5.4

CAL Spec - Calibration wavelengths

Tab: Method

CAL Spec Properties... Calibration wavelengths

Command name

Name of the command.

Input

25 characters

The reference wavelengths with which the spectrometer is calibrated can
be entered on this tab.

Tolerance

Tolerance for the preset wavelength when assigning the highest peak
maximum found during the calibration measurement to the reference
wavelength. If a peak is outside the range Wavelength +/- Tolerance,
the calibration will be canceled. The existing calibration in the configura-
tion remains in effect.

Range

0.1 ... 1100.0 nm

Default value

0.5 nm

[New]

Open the dialog window Calibration wavelength # to enter wave-
lengths and tolerances (see Chapter 5.6.5.5.5, page 1107).

[Properties]

Open the dialog window Calibration wavelength # to edit the values
selected in the table (see Chapter 5.6.5.5.5, page 1107).

[Delete]

Delete the selected line.

5.6.5.5.5

CAL Spec - Calibration wavelength #

Tab: Method

CAL Spec Properties... Calibration wavelengths [New]/

[Properties]

Calibration wavelength #

Wavelength

Range

200.0 ... 1100.0 nm

Default value

'empty' nm

Advertising