7 properties - analog outputs, Properties - analog outputs 8 – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1634

7.20 IO controller
■■■■■■■■■■■■■■■■■■■■■■
1618
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Note
With the aid of the following parameters the physical values that are
applied to the analog input can be mapped to any measured value.
The transfer function is linear.
0% Mapping
Measured value to be used when the Low level (0%) is applied as input
value.
Range
-214783648 ... 214783647
Default value
0
100% Mapping
Measured value to be used when the High level (100%) is applied as
input value.
Range
-214783648 ... 214783647
Default value
100
Unit
Unit of the measured value to be used.
Input
50 characters
Default value
%
7.20.7
Properties - Analog outputs
On the tab Analog outputs all outputs detected are automatically listed
in a table. For each analog output the settings can be modified.
Port
Display of the internal description of the port.
Port description
Display of the port description.
Port specification
Shows the specifications of the port.
Type
Display of the port type.
Low level (0%)
Display of the lowest analog output value with unit.