5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1099
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.
Stirrer
This section is visible only if the 856 Conductivity Module was selected as
device type.
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: off)
If this option is enabled, the stirrer will be switched off automatically
when the command has finished.
5.6.5.4.3
CAL Cond - Measuring parameters
Tab: Method
▶ CAL Cond ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement
Measurement frequency
712
Selection
auto | 300 Hz | 2.4 kHz
Default value
auto