8.4 Manual control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1686
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Status
Display of the current status.
Selection
ACTIVE | INACTIVE | ERROR
[Start]
Sets the digital output to ACTIVE.
[Stop]
Sets the digital output to INACTIVE.
Timer control
Port description
Display of the port description.
Port
Display of the internal description of the port.
Type
Display of the port type.
Parameter 1
The meaning of Parameter 1 depends on the Type of the digital output
that has been set.
When Type = Stirrer: stirrer output in %, pulse duration dependent on
Parameter 2.
Range
-214783647 ... 214783647
Default value
100
Parameter 2
The meaning of Parameter 2 depends on the Type of the digital output
that has been set.
For Type = Stirrer:
■
0: Duration of the switch-on pulse fixed at 500 ms.
■
>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in
10 ms
Example:
Parameter 1 = 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on
225 ms, switched off 75 ms
Range
-214783647 ... 214783647
Default value
0