8.4 Manual control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1694
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Type
Display of the port type.
Distance
Distance to be moved to.
Visible only in cases of stepping motors where Type = Linear drive.
Range
-214783647 ... 214783647 mm
Default value
0 mm
Angle
The angle to be moved to.
Not visible in cases of stepping motors where Type = Rotary drive.
Range
-214783647 ... 214783647 °
Default value
0 °
Timeout
Time in seconds or minutes within which the action must have run
through completely. If this is not done after the maximum waiting time
has expired, the status of the stepping motor will be set to ERROR.
Range
0 ... 999999
Default value
10
Selection
s | min
Default value
s
Status
Display of the current status.
Selection
ACTIVE | INACTIVE | ERROR
[Start]
Sets the stepping motor to ACTIVE.
[Stop]
Sets the stepping motor to INACTIVE.
End position
On this tab the position of the stepping motor can be modified manually,
it can be moved to the maximum end position.
Port description
Display of the port description.