5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1185

command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Tandem dosing

on | off (Default value: off)

If this option is enabled, then dosing will proceed continuously with two
dosing devices in combination. The second dosing device is dosing while
the first one is being filled and vice-versa. This parameter is displayed only
for Titrandos and 855.

Dosing device

Selection of the second dosing device (exchange or dosing unit) to be
used for the dosing if the first dosing device is not available.

Titrando

Selection

1 | 2 | 3 | 4

Default value

2

855

Selection

1 | 2 | 3

Default value

2

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked.

Titrando, 855
Selection

Solution name | not defined

Default value

not defined

not defined
No tests will be carried out.

Filling rate

Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the