2 elt loop - properties – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1134

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1118
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Desig-
nation
Description
.ETR
Result of the electrode test:
0 = No data available
1 = Short circuit
2 = Excessive start drift
3 = Wrong buffer
4 = Problems with the diaphragm
5 = Reference system defective
6 = No buffer values for the temperature or excessive devia-
tion between the buffers
7 = Partial short circuit
8 = Glass defective
9 = Poor electrode
10 = Usable electrode
11 = Good electrode
12 = Excellent electrode
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command has
never ended
.LCO
Loop counter = current quantity of completed runs, both of
Repeat loops and of While loops
.LST
Start time of the loop command (Date/Time)
.SLO
Electrode slope of the sensor used for the command (in mV
for ISE sensors)
5.6.5.8.2
ELT LOOP - Properties
Dialog window: Method
▶ ELT LOOP ▶ Properties...
Command name
Name of the command.
Input
25 characters
In this dialog window, the general parameters are set to the buffer used
and to the electrode test.