5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1233
Selection
774 Oven Sample Processor | 874 Oven Sam-
ple Processor
Default value
874 Oven Sample Processor
Gas
Gas type
Gas type of the carrier gas.
Selection
Air | Nitrogen | other gas
Default value
Air
Gas flow factor
Factor for the correct determination of the gas flow. This parameter is
only visible for Gas type = other gas. The values for frequently used car-
rier gases are listed in the device manuals.
Range
0.001 ... 9.999
Default value
1.000
Gas flow
Inlet
Selection of the inlet of the carrier gas
Selection
Pump | Valve
Default value
Pump
Pump
Use of ambient air as carrier gas.
Valve
Use of gas out of a compressed gas cylinder.
Flow rate
874
Range
10 ... 150 mL/min
Default value
50 mL/min
Gas flow
Selection
Switch on | Switch off
Default value
Switch on
Switch on
Switches on the gas flow.
Switch off
Switches off the gas flow.