Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 830

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
814
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Range
0 ... 999999 s
Default value
1 s
Pause 1
Pause 1
Waiting time, e.g. for the electrode to settle down before a start volume is
added.
Range
0 ... 999999 s
Default value
0 s
Start volume
Start volume
Volume to be dosed with the indicated dosing rate before the start of the
titration.
Titrando
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Titrino
Range
0.00 ... 999.99 mL
Default value
0.00 mL
Dosing rate
Rate at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.
Titrando
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum