5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
943
Titrando, 855, 867
Range
0.1 ... 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Range
0.08 ... 16200 s (Increment: 0.08)
Default value
2.0 s
Stop measured value pH
Stops when the preset measured value has been reached since the start of
the measurement.
Titrando, 855, 867
Range
-20.000 ... 20.000
Selection
off
Default value
off
off
No stop.
Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:
Measuring time
Maximum period of time for measurement.
Titrando, 855, 867
Range
0 ... 999999 s
Default value
120 s
Titrino
Range
0 ... 9999 s
Default value
120 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Titrando, 855, 867
Range
0.1 ... 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Range
0.08 ... 16200 s (Increment: 0.08)
Default value
2.0 s