5 dos ph - monitoring, Monitoring – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1191

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1175
Selection
off
off
Means that no stop will take place.
Stop time
Stops when the preset time has elapsed since the start of the dosing.
Range
0 ... 999999 s
Selection
off
Default value
off
off
Means that no stop will take place.
Filling rate
Rate at which the dosing cylinder is to be filled after dosing. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used. If the entered filling rate is too high for the selected
dosing device, it will automatically be reduced during dosing to the largest
possible value.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min
Note
The filling rate should be reduced for viscous liquids.
5.6.6.3.4.5
DOS pH - Monitoring
Tab: Method
▶ DOS pH ▶ Properties... ▶ Monitoring
Command name
Name of the command.
Input
25 characters
The following monitoring processes can be switched on and defined on
this tab: