3 meas spec - measuring parameters – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1055

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1039

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead.

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Avantes Spectrometer

Default value

Avantes Spectrometer

5.6.4.13.3

MEAS Spec - Measuring parameters

Tab: Method

MEAS Spec Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Start wavelength

Lower limit of the spectrum.

Range

100.0 ... 2000.0 nm

Default value

400.0 nm

End wavelength

Upper limit of the spectrum.

Range

100.0 ... 2000.0 nm

Default value

1000.0 nm

Integration time

Integration time for spectrum recording.

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.

Range

1 ... 10000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.