Range
1 ... 9
Default value
2
Dosing rate
Rate at which the solution is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit used (see below).
When the function is carried out the rate will automatically be decreased
to the highest possible value.
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Filling rate
Rate at which the cylinder is to be filled. The maximum filling rate depends
on the cylinder volume of the exchange unit used (see below). When the
function is carried out the rate will automatically be decreased to the
highest possible value.
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Maximum dosing/filling rate for the exchange unit, depending on
the cylinder volume:
Cylinder volume
maximum rate
1 mL
3.0 mL/min
5 mL
15.0 mL/min
10 mL
30.0 mL/min
20 mL
60.0 mL/min
50 mL
150.0 mL/min
Note
Enter lower rates for high-viscosity liquids.
Tubing parameters
Definition of the length and diameter of the tubing connected to the
exchange unit.