5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
669
Window = Measured value
Range
-200.0 ... 200.0 µA
Default value
200.0 µA
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
Window = Time
Range
0 ... 999999 s
Default value
999999 s
Gran evaluation
Note
This evaluation method is possible only with Titrando and 855.
Gran evaluation
on | off (Default value: off)
If this option is enabled, then titration curves will be evaluated in accord-
ance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 930).
Initial volume
Volume which is present in the measuring vessel before the command is
started.
Range
0.01 ... 9999.99 mL
Default value
50.00 mL
Lower limit
Value for the lower limit of the Gran evaluation range.
Range
-200.0 ... 200.0 µA
Default value
-200.0 µA
Upper limit
Value for the upper limit of the Gran evaluation range.
Range
-200.0 ... 200.0 µA
Default value
200.0 µA