5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
969
Stop measured value
Stops when the preset measured value has been reached since the start of
the measurement.
Titrando, 855, 867
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off
off
No stop.
Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:
Measuring time
Maximum period of time for measurement.
Titrando, 855, 867
Range
0 ... 999999 s
Default value
120 s
Titrino
Range
0 ... 9999 s
Default value
120 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Titrando, 855, 867
Range
0.1 ... 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Range
0.08 ... 16200 s (Increment: 0.08)
Default value
2.0 s
Stop measured value
Stops when the preset measured value has been reached since the start of
the measurement.
Titrando, 855, 867
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off