5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

898

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Range

0 ... 999999 s

Default value

1 s

Pause 1

Pause 1

Waiting time, e.g. for the electrode to settle down before a start volume is
added.

Titrando, 855
Range

0 ... 999999 s

Default value

0 s

Start volume

Start volume

Volume to be added before the start of the titration at the dosing rate
indicated.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Titrino
Range

0.00 ... 999.99 mL

Default value

0.00 mL

Dosing rate

Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum