2.4 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Designation
Description
Commands
.ETE
End temperature (temperature fol-
lowing processing of the com-
mand) in °C
DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without T/Flow,
Ref, Spec), CAL
Cond, CAL MEAS,
STDADD, DOS,
ELT MEAS
.ETR
Result of the electrode test.
Electrode quality:
0 = No data available
1 = Short circuit
2 = Excessive start drift
3 = Wrong buffer
4 = Problems with the diaphragm
5 = Reference system defective
6 = No buffer values for the tem-
perature or excessive deviation of
the temperature between buffers
7 = Partial short circuit
8 = Glass defective
9 = Poor electrode
10 = Usable electrode
11 = Good electrode
12 = Excellent electrode
ELT LOOP
.EVT
End volume (total dosed volume
at the end of the command) in mL
DET, MET, SET,
KFT, STAT, DOS
.FIN
Command status; 1 = Command
has ended at least once, 0,
invalid (variable not available) =
Command has never ended
All except START
and END
.FP{x}.CNT
Intensity for the fixed endpoint x
(1…9) in counts
MEAS Opt
.FP{x}.DME
Measured value differential for the
fixed endpoint x (1…9)
MET