5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

884

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Titrando, 855
Range

0.0 ... 150.0 °C

Default value

0.2 °C

Upper limit

Upper limit of the temperature. If the temperature exceeds this limit, the
event Temperature upper limit exceeded is triggered.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

150.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

500.0 °C

Upper hysteresis

Upper hysteresis of the temperature. If the temperature once again falls
below the upper limit by this hysteresis value, then the event Tempera-
ture upper limit ok
is triggered.

Titrando, 855
Range

0.0 ... 150.0 °C

Default value

0.2 °C

Action

Selection of the event to be triggered if the upper or lower limit is excee-
ded:

Selection

Cancel determination | Cancel command
(Titrando only)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando only)
The running STAT command will be canceled, then the next com-
mand will be carried out.
Wait for [Continue]
Reagent dosing in the running STAT command will be interrupted and
a message will be displayed. As soon as the monitored measured value
is again within the limits (including hysteresis) reagent dosing can be
resumed by pressing the [Continue] button in this message box.