5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1032
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
5.6.4.11.3
MEAS TC Cond - Measuring parameters
Tab: Method
▶ MEAS TC Cond ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement frequency
712
Selection
auto | 300 Hz | 2.4 kHz
Default value
auto
auto
The more suitable of the two frequencies is selected.
Stop time
Maximum period of time for measurement.
Range
0 ... 999999 s
Default value
3000 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
712
Range
0.4 ... 999999 s
Default value
2.0 s
856
Range
0.1 ... 999999 s
Default value
2.0 s
Sample solution
Selection of the sample solution from the table Sample solutions (TC
conductivity).
Selection Table entries
Default value first entry of table
Start temperature
Temperature at the start of the measurement.
Range
-20.0 ... 150.0 °C
Default value
20.0 °C