5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

487

Fix value

Value of the fix endpoint.

Measured value
Range

-20.000 ... 20.000 pH

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.2.3

DET U

5.6.3.2.3.1

DET U - Overview

Dialog window: Method

▶ DET U ▶ Properties... ▶ DET U - 'Command name'

Command for Dynamic equivalence point titrations with potentio-
metric voltage measurement.

Instruments
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the command DET U are configured in the following
7 tabs:

■

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

■

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

■

Titration parameters
Parameters for the run of the titration.