8.4 Manual control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1676
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
[Stop]
Stop the heater.
Gas flow
Inlet
Selection of the inlet of the carrier gas.
Selection
Pump | Valve
Default value
Pump
Pump
Use of ambient air as carrier gas.
Valve
Use of gas out of a compressed gas cylinder.
Flow rate
874
Range
10 ... 150 mL/min
Default value
50 mL/min
Gas type
Gas type of the carrier gas.
Selection
Air | Nitrogen | Other gas
Default value
Air
Gas flow factor
Factor for the correct determination of the gas flow. This parameter is
only displayed for Gas type = Other gas. The values for frequently used
carrier gases are listed in the manual of the 774 and 874 Oven Sample
Processor.
Range
0.001 ... 9.999
Default value
1.000
[Start]
Switch on the gas flow. In the status display the running time since start is
shown.
[Stop]
Switch off the gas flow.
Gas flow pump
Switch the gas flow pump on the Oven Sample Processor on/off.