3 kfc - start conditions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 849

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
833
Note
For generator electrodes of the type with diaphragm, it is recommen-
ded to set the Generator current to auto.
For generator electrodes without diaphragm it is recommended to
set the Generator current to 400 mA.
Stirrer
Stirrer
Selection of the stirrer.
851, 852
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
851, 852
Range
-15 ... 15
Default value
8
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command.
5.6.3.6.3
KFC - Start conditions
Tab: Method
▶ KFC ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
Pause
Pause
Waiting time before the start of the titration.