Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 518

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
502
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Stop measured value
Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off
Titrino
Range
-2000 ... 2000 mV
Selection
off
Default value
off
Stop EP
The titration is stopped when the specified number of equivalence points
has been found.
Range
1 ... 9
Default value
9
Selection
off
Volume after EP
When the number of equivalence points defined under Stop EP has been
found, this volume will be added. In this way you can see the curve shape
after the equivalence point is reached.
Titrando, 855
Range
0.01000 ... 9999.99 mL
Selection
off
Default value
off
Stop time
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.
Titrando, 855
Range
0 ... 999999 s
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.