Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual
Page 1047

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1 (for Process Analysis)
■■■■■■■■
1031
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display of the device type.
Selection
712 Conductometer | 856 Conductivity Mod-
ule
Default value
856 Conductivity Module
Sensor
Sensor
Selection of a sensor of the Conductivity sensor type from the sensors
configured in the sensor table.
Selection
Sensor name | Conductivity sensor
Default value
Conductivity sensor
Stirrer
This section is visible only if the 856 Conductivity Module was selected as
device type.
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished.