5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

608

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Volume increment

Small volume increments are used for the determination of blank values or
for very asymmetrical curves. A good guideline is 1/20 of the expected EP
volume. For steep jumps the volume increment should tend toward 1/100
and for flat jumps toward 1/10 of the EP volume. Small volume incre-
ments are used for determining blank values or with very asymmetrical
curves. The accuracy of the evaluation cannot be increased by using
smaller increments as the measured value alterations between two meas-
uring points are then of the same order of magnitude as the noise.

Titrando, 855
Range

0.0001 ... 9.9999 mL

Default value

0.05 (slow), 0.10 (optimal), 0.20 (fast) mL

Titrino
Range

0.001 ... 9.999 mL

Default value

0.05 (slow), 0.10 (optimal), 0.20 (fast) mL

Dosing rate

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware
tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously. This value is used for temperature
correction in pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C