3  Programming the Analyzer

IEEE 488.2 Common Commands

IEEE 488.2 Common Commands

The instrument supports the following subset of IEEE 488.2 Common Commands, as defined in
Chapter 10 of

IEEE Standard 488.2–1992

. As indicated below, some of these commands

correspond directly to instrument front-panel key functionality, while others are available only as
remote commands.

Commands

Description

*CAL?

"All" on page 1269

*CLS

"Clear Status " on page 103

*ESE <int>

*ESE?

"Standard Event Status Enable " on page 104

*ESR?

"Standard Event Status Register Query " on page 105

*IDN?

"Identification Query " on page 105

*OPC

*OPC?

"Operation Complete " on page 106

*OPT?

"Query Instrument Options " on page 107

*RCL <reg>

"Recall Instrument State " on page 107

*RST

"*RST (Remote Command Only)" on page 108

*SAV <reg>

"Save Instrument State " on page 108

*SRE <int>

*SRE?

"Service Request Enable " on page 109

*STB?

"Status Byte Query " on page 109

*TRG

"Trigger  " on page 109

*TST?

"Self Test Query " on page 110

*WAI

"Wait-to-Continue " on page 110

All

Immediately executes an alignment of all subsystems. The instrument stops any measurement
currently underway, performs the alignment, then restarts the measurement from the beginning
(similar to pressing the Restart key).

If an interfering user signal is present at the RF Input, the alignment is performed on all
subsystems except the RF. After completion, the Error Condition message “Align skipped: 50 MHz
interference” or “Align skipped: 4.8 GHz interference” is generated. In addition the Error

Remote Language Compatibility Measurement Application Reference

101