6  RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale

Example

POW:ATT:STEP 2

Notes

Note this feature works like a 1-N choice from the front panel, but it takes a specific value
(in dB) when used remotely. The only valid values are 2 and 10.

Dependencies

Blanked in CXA and EXA if option FSA (2 dB steps) is not present. If blanked, attempts to set it
via SCPI will yield an error.    

Couplings

When the attenuation step size changes, the current mechanical attenuation value is adjusted
(if necessary) to be quantized to the new step size. That is, if step is set to 10 dB, mech atten is
increased if necessary so it is a multiple of 10 dB

Preset

PXA and MXA: 2 dB

EXA and CXA: 10 dB (2 dB with option FSA)

State Saved

Saved in instrument state

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.03.00

Max Mixer Level

Controls the limitation on the Ref Level for a given attenuation setting, and therefore also
interacts with the Auto rules for selecting the attenuation as a coupling from the reference level.

Key Path

AMPTD Y Scale, Attenuation

Remote Command

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <real>

[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?

Example

POW:MIX:RANG –15 dBm

Preset

–10 dBm

State Saved

Saved in instrument state

Min

–50 dBm

Max

–10 dBm

Default Unit

Depends on the current selected Y axis unit, see Swept SA discussion of Y Axis Unit

Initial S/W Revision

Prior to A.02.00

528

Remote Language Compatibility Measurement Application Reference