6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Source

[None] on a Restore Input/Output Defaults.

If an internal Tracking Generator is installed, then instead of  None, the default selection will be
INTERNAL.

State Saved

Selected Source is

l

Power On Persistent (survives power cycle)

l

Part of the Input/Output system, which means it is Loaded and Saved with state.

Readback

Two lines of readback give the type information and serial number of the current source on the
Select Source key in the form

[<source type>]
[<serial number>]

[None] shows in the type area and blank in the serial number area if a source has not been
configured.

[Internal TG] shows in type area and serial number in the serial number area if an internal
Tracking Generator has been selected. 

For example: [MXG]/n, [US01020022].  This indicates an MXG of serial number US01020022.

Backwards
Compatibility SCPI

:SYSTem:COMMunicate:LAN:SOURce[:EXTernal]:IP <address string>

:SYSTem:COMMunicate:LAN:SOURce[:EXTernal]:IP?

This command is provided for compatibility with PSA Option 215.  The address string is
reformatted for the X-Series. For example, if the customer sends

:SYSTem:COMMunicate:LAN:SOURce:EXTernal:IP 146.208.172.111

The analyzer turns this into

:SYSTem:COMMunicate:SOURce:ADDRess “TCPIP0::146.208.172.111::INSTR"

Initial S/W Revision

A.06.01

Modified at S/W
Revision

A.10.01

Source Preset

The Source Preset key forces all the settings in the analyzer’s Source State to their preset
condition. 

The Source State is the set of Source settings that is maintained and remembered by the analyzer
for use in the Tracking Source Mode.  The Source State variables are controlled and set in the
menus under the Source front panel key.  These settings include:

l

RF Output Off

l

Amplitude = – 10 dBm

Remote Language Compatibility Measurement Application Reference