6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Sweep/Control

Preset

DRANge

State Saved

Saved in instrument state

Backwards Compatibility
Notes

The legacy parameter DYNamicrange is unsupported

Initial S/W Revision

Prior to A.02.00

Auto

This selection is automatically chosen when Auto Couple is pressed. When in Auto, the Sweep
Type Rules are set to Best Dynamic Range. It seems like a very simple Auto function but the use
of this construct allows a consistent statement about what the Auto Couple key does.

Key Path

Sweep/Control, Sweep Setup, Sweep Type Rules

Remote Command

[:SENSe]:SWEep:TYPE:AUTO:RULes:AUTO[:STATe] OFF|ON|0|1

[:SENSe]:SWEep:TYPE:AUTO:RULes:AUTO[:STATe]?

Example

:SWE:TYPE:AUTO:RUL:AUTO ON

Couplings

Pressing Auto Couple always sets Sweep Type Rules to Auto.

Preset

ON

State Saved

Saved in instrument state

Initial S/W Revision

Prior to A.02.00

Best Dynamic Range

This selection tells the analyzer to choose between swept and FFT analysis with the primary goal
of optimizing dynamic range. If the dynamic range is very close between swept and FFT, then it
chooses the faster one. This auto selection also depends on RBW Type.

In determining the Swept or FFT setting, the auto rules use the following approach:

If the RBW Filter Type is Gaussian use the RBW for the Normal Filter BW and if that RBW >

210 Hz, use swept; for RBW <= 210 Hz, use FFT

If the RBW Filter Type is Flat Top, use the same algorithm but use 420 Hz instead of 210 Hz for

the transition point between Swept and FFT

If any of the CISPR detectors is chosen for any active trace, always use Swept.

Key Path

Sweep/Control, Sweep Setup, Sweep Type Rules

Remote Language Compatibility Measurement Application Reference