6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

Key Path

Input/Output, External Gain

Remote Command

[:SENSe]:CORRection:SA[:RF]:GAIN <rel_ampl>

[:SENSe]:CORRection:SA[:RF]:GAIN?

Example

CORR:SA:GAIN 10 sets the Ext Gain value to 10 dB

CORR:SA:GAIN –10 sets the Ext Gain value to –10 dB (that is, an attenuation of 10 dB)

Notes

Does not auto return.

Dependencies

The reference level limits are determined in part by the External Gain/Atten, Max Mixer Level,
and RF Atten.

This key is grayed out in Modes that do not support External Gain

Preset

This is unaffected by Preset but is set to 0 dB on a "Restore Input/Output Defaults" or "Restore
System Defaults->All"

State Saved

Saved in instrument state

Min

–120 dB

Max

120 dB

Readback

Preamp Gain, <Ext Gain value> dB

Backwards
Compatibility SCPI

[:SENSe]:CORRection:OFFSet[:MAGNitude]

The legacy "Ext Preamp Gain" key is now called "Ext Gain" and the sub-menu has choices of
Ext Preamp | MS | BTS for backwards compatibility.

Initial S/W Revision

Prior to A.02.00

MS

Sets an external gain/attenuation value for MS (Mobile Station) tests.

Key Path

Input/Output, External Gain

Remote Command

[:SENSe]:CORRection:MS[:RF]:GAIN <rel_ampl>

[:SENSe]:CORRection:MS[:RF]:GAIN?

Example

CORR:MS:GAIN 10 sets the Ext Gain value to 10 dB

CORR:MS:GAIN –10 sets the Ext Gain value to –10 dB (that is, a loss of 10 dB.)

Notes

Does not auto return.

Dependencies

The reference level limits are determined in part by the External Gain, Max Mixer Level, RF Atten

This key is grayed out in modes that do not support MS.

Preset

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or
"Restore System Defaults->All"

620

Remote Language Compatibility Measurement Application Reference