Agilent Technologies Signal Analyzer N9030a User Manual

Page 622

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output

This key is grayed out in modes that do not support BTS.

Preset

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or
"Restore System Defaults->All"

State Saved

Saved in instrument state.

Min

–100 dB

Max

100 dB

Readback

BTS, <Ext Gain value> dB

Initial S/W Revision

Prior to A.02.00

Remote Command

[:SENSe]:CORRection:BTS[:RF]:LOSS <rel_ampl>

[:SENSe]:CORRection:BTS[:RF]:LOSS?

Example

CORR:BTS:LOSS 10  sets the Ext Gain value to –10 dB, and subsequently querying :LOSS will
give 10 dB

CORR:BTS:LOSS –10 sets the Ext Gain value to 10 dB, and subsequently querying :LOSS will give
–10 dB

Notes

A positive value of <rel_ampl> in the above command means a loss and a negative value
indicates a gain.

Anytime :LOSS is set it sets :GAIN to the negative value of the parameter sent.  

Anytime :LOSS is queried it gives the negative of :GAIN

Preset

This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or
"Restore System Defaults->All"

Min

100 dB

Max

–100 dB

Initial S/W Revision

Prior to A.02.00

I Ext Gain

This function affects only the I channel input, except when the Input Path is I+jQ.  In I+jQ this
setting is applied to both I and Q channel inputs. It is not available unless the Baseband I/Q
option (BBA) is installed.

Key Path

Input/Output, External Gain

Remote Command

[:SENSe]:CORRection:IQ:I:GAIN <rel_ampl>

[:SENSe]:CORRection:IQ:I:GAIN?

622