Agilent Technologies Signal Analyzer N9030a User Manual

Page 1247

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

System

:SYSTem:ERRor:OVERload OFF command.  In either case, Input Overloads always set the status
bit.

For versions of firmware before A.10.01, the Input Overload was only a Warning and so was never
available in the SCPI queue, although it did set the status bit.  For A.10.01 and later, the Input Overload is
an error and can be enabled to the SCPI queue using this command.

Remote Command

:SYSTem:ERRor:OVERload[:STATe] 0|1|OFF|ON

Example

:SYST:ERR:OVER 1  Enable overload errors

Preset

Set to OFF by Restore Misc Defaults (no Overload errors go to SCPI)

State Saved

Saved in instrument state.

Initial S/W Revision

A.10.01

System

The System screen is formatted into three groupings: product descriptive information, options
tied to the hardware, and software products:

The Previous Page is grayed-out if the first page of information is presently displayed. The Next
Page menu key is grayed-out if the last page is information is presently displayed.

Remote Language Compatibility Measurement Application Reference