6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Mode Setup

Restore Mode Defaults

Resets the state for the currently active mode by resetting the mode persistent setting to their
default values and by performing a mode preset. This function never causes a mode switch. This
function performs a full preset on the active mode.

Remote Command

:INSTrument:DEFault

Key Path

Mode Setup

Preset Type (Remote Command Only)

For compatibility with ESA/PSA, the command PRESet:TYPE is implemented as a no-op.

Mode

All

Remote Command

:SYSTem:PRESet:TYPE FACTory|MODE|USER

:SYSTem:PRESet:TYPE?

Example

:SYST:PRES:TYPE FACT

Notes

This command is supported for backward compatibility only.  It is a no-op and does not change
the behavior of any preset operation.

Preset

This is unaffected by Preset, but is set to MODE by Restore System Defaults, All.

State Saved

No

Initial S/W Revision

Prior to A.02.00

Global Settings

Opens a menu that allows you to switch certain Meas Global parameters to a Mode Global state.
These switches apply to all Modes that support global settings.  No matter what Mode you are in
when you set the “Global Center Frequency” switch to on, it applies to all Modes that support
Global Settings.

Key Path

Mode Setup

Initial S/W Revision

Prior to A.02.00

Global Center Freq

The software maintains a Mode Global value called “Global Center Freq”. 

892

Remote Language Compatibility Measurement Application Reference