6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup

High (Best Log Accy)

When ADC dither is set to High, the scale fidelity is especially good, most notably the relative
scale fidelity. The tradeoff is that there is a modest loss of noise floor performance, up to about a
decibel.

Key Path

Meas setup, ADC Dither

Example

ADC:DITH:HIGH

Readback

If manually selected, the readback is High, with the “Man” underlined

Initial S/W Revision

A.02.00

Medium (Log Accy)

The Medium setting of ADC Dither (known as “On” in earlier versions of the instrument software)
improves the  linearity of low-level signals at the expense of some noise degradation.

Key Path

Meas setup, ADC Dither

Example

ADC:DITH:ON

Readback

If manually selected, the readback is Medium, with the “Man” underlined

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.02.00

Off (Best Noise)

When ADC Dither is Off, the instrument noise floor is improved, because without the need to
make room for the dither, you get a lower noise floor and better sensitivity.

Key Path

Meas setup, ADC Dither

Example

ADC:DITH:OFF

Readback

If manually selected, the readback is Off, with the “Man” underlined.

Initial S/W Revision

Prior to A.02.00

Swept IF Gain

To take full advantage of the RF dynamic range of the analyzer, there is an added switched IF
amplifier with approximately 10 dB of gain. When you can turn it on without overloading the

840

Remote Language Compatibility Measurement Application Reference