3  Programming the Analyzer
STATus Subsystem

Questionable Integrity Signal Event Query

This query returns the decimal value of the sum of the bits in the Questionable Integrity Signal
Event register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:QUEStionable:INTegrity:SIGNal[:EVENt]?

Example

STAT:QUES:INT:SIGN?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Integrity Signal Negative Transition

This command determines which bits in the Questionable Integrity Signal Condition register will
set the corresponding bit in the Questionable Integrity Signal Event register when the condition
register bit has a negative transition (1 to 0). The variable <integer> is the sum of the decimal
values of the bits that you want to enable.

Mode

All

Remote Command

:STATus:QUEStionable:INTegrity:SIGNal:NTRansition <integer>

:STATus:QUEStionable:INTegrity:SIGNal:NTRansition?

Example

STAT:QUES:INT:SIGN:NTR 4

Burst found will be reported to the Integrity Summary of the Status Questionable register.

Preset

0

Min

0

Max

32767

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Integrity Signal Positive Transition

This command determines which bits in the Questionable Integrity Signal Condition register will
set the corresponding bit in the Questionable Integrity Signal Event register when the condition

164