6  RLC Swept SA Measurement Front-Panel & SCPI Reference

BW

Key Path

BW, RBW Control, Filter BW

Example

BAND:TYPE DB3

Readback

–3 dB

Initial S/W Revision

Prior to A.02.00

–6 dB

Selects the filter bandwidths where the bandwidth is defined at the –6 dB points. This uses the
normal RBW filters, but the value displayed on the key, active function line and screen
annotation changes to reflect the –6 dB bandwidth instead of the –3 dB bandwidth.

Key Path

BW, RBW Control, Filter BW

Example

BAND:TYPE DB6

Readback

–6 dB

Initial S/W Revision

Prior to A.02.00

Noise

Selects the noise filter bandwidths. This uses the normal RBW filters, but the value displayed on
the key, active function line and screen annotation changes to reflect the equivalent noise
bandwidth, instead of the –3 dB bandwidth.

Key Path

BW, RBW Control, Filter BW

Example

BAND:TYPE NOIS

Readback

Noise

Initial S/W Revision

Prior to A.02.00

Impulse

Selects the impulse bandwidths. This uses the normal RBW filters, but the value displayed on the
key, active function line and screen annotation changes to reflect the equivalent impulse
bandwidth instead of the –3 dB bandwidth.

Key Path

BW, RBW Control, Filter BW

Example

BAND:TYPE IMP

Readback

Impulse

Initial S/W Revision

Prior to A.02.00