Agilent Technologies Signal Analyzer N9030a User Manual

Page 171

Advertising
background image

3  Programming the Analyzer

STATus Subsystem

Max

32767

Status
Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Questionable Temperature Register

The following commands and queries are available for this register:

"Questionable Temperature Condition" on page 171

"Questionable Temperature Enable" on page 171

"Questionable Temperature Event Query" on page 172

"Questionable Temperature Negative Transition" on page 172

"Questionable Temperature Positive Transition" on page 173

Questionable Temperature Condition

This query returns the decimal value of the sum of the bits in the Questionable Temperature
Condition register.

The data in this register is continuously updated and reflects the current conditions.

Mode

All

Remote Command

:STATus:QUEStionable:TEMPerature:CONDition?

Example

STAT:QUES:TEMP:COND?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Temperature Enable

This command determines which bits in the Questionable Temperature Condition Register will set
bits in the Questionable Temperature Event register, which also sets the Temperature Summary bit
(bit 4) in the Questionable Register.

The variable <integer> is the sum of the decimal values of the bits you want to enable.

Remote Language Compatibility Measurement Application Reference